E1

Omron Blood Pressure Monitor

Severity: Minor

What it means

Omron's instruction manuals for the M3 (HEM-7154-E) and M3 Comfort (HEM-7155-E) put E1 under the problem "E1 appears or the arm cuff does not inflate".
They give four causes: START/STOP was pressed while the cuff wasn't on your arm, the air plug isn't fully plugged into the monitor, the cuff isn't applied correctly, or air is leaking from the cuff.
Other Omron monitors may use E1 for something else, so check the table in your own monitor's manual.

Affected Models

  • Omron M3 (HEM-7154-E), M3 Comfort (HEM-7155-E) and X3 Comfort (HEM-7155-EO) upper arm monitors

How to Fix It

  1. Press START/STOP to turn the monitor off.

  2. Push the air plug firmly into the monitor.

  3. Put the arm cuff on correctly, then press START/STOP to measure again.

    Section 5 of the instruction manual shows how to apply the cuff.

  4. If air keeps leaking from the cuff, replace the cuff with a new one.

    Section 12 of the manual covers replacing it.