System fault, refer to user guide, Error 004
ResMed CPAP Machine
Severity: ModerateWhat it means
ResMed's support pages list System fault, refer to user guide, Error 004 Description: Main PCB over temperature on the AirSense 10, AirStart, AirCurve and Lumis devices.
The device's main circuit board has become too hot.
ResMed's possible causes: the room conditions, a blocked filter, blocked air tubing, water in the air tubing, or a faulty main board.
How to Fix It
-
Make sure the device is not covered, and that the room temperature is within the device's operating range.
-
Let the device cool down, then disconnect the power supply and reconnect it to restart the device.
-
Replace the air filter.
-
Check the air tubing and remove any blockages, then press the dial to clear the message.
-
Empty any water out of the air tubing, then disconnect and reconnect the power supply.
When to Call a Professional
If none of this clears it, ResMed's last step is replacing the main circuit board: contact your equipment provider, the company that supplied the machine.