System fault, refer to user guide, Error 004

ResMed CPAP Machine

Severity: Moderate

What it means

ResMed's support pages list System fault, refer to user guide, Error 004 Description: Main PCB over temperature on the AirSense 10, AirStart, AirCurve and Lumis devices.
The device's main circuit board has become too hot.
ResMed's possible causes: the room conditions, a blocked filter, blocked air tubing, water in the air tubing, or a faulty main board.

How to Fix It

  1. Make sure the device is not covered, and that the room temperature is within the device's operating range.

  2. Let the device cool down, then disconnect the power supply and reconnect it to restart the device.

  3. Replace the air filter.

  4. Check the air tubing and remove any blockages, then press the dial to clear the message.

  5. Empty any water out of the air tubing, then disconnect and reconnect the power supply.

When to Call a Professional

If none of this clears it, ResMed's last step is replacing the main circuit board: contact your equipment provider, the company that supplied the machine.