Error 4
OneTouch Blood Glucose Meter
Severity: MinorWhat it means
OneTouch's support pages say one of these may apply with Error 4:
Not enough blood or control solution was applied, or more was added after the meter began to count down.
The test strip may have been damaged or moved during testing.
The sample was applied improperly, or there may be a problem with the meter.
How to Fix It
-
Repeat the test with a new test strip.
-
If the error message appears again, contact OneTouch Customer Care.
Frequently Asked Questions
My Verio Reflect shows Error 4 when I test with control solution. Why?
OneTouch says Error 4, sub-code 2 can come from testing with control solution without setting the meter to show Apply Control Solution.
Go into the meter Settings and select Control Solution, insert the test strip, and wait for Apply Control Solution to appear.
Why does my Ultra 2 show Error 4 in the cold?
On the Ultra 2, OneTouch says Error 4 can mean you have high glucose and tested near the low end of the operating temperature range, 43 to 111 °F (about 6 to 44 °C).
If you tested somewhere cool, repeat the test somewhere warmer with a new test strip.