Err

Braun Thermometer

Severity: Minor

What it means

Braun's FAQs say the Err message appears when the room temperature is outside the thermometer's operating range of 10 to 40 °C (50 to 104 °F).
The same answer is on Braun's FAQ pages for the ThermoScan 3, 5, 6, 7 and 7+ ear thermometers, the No Touch, forehead and TempleSwipe thermometers, and the digital stick thermometer.

How to Fix It

  1. Leave the thermometer for 30 minutes in a room between 10 and 40 °C (50 and 104 °F).

    On the Forehead Thermometer (BFH175), Braun also says to press the power button to reset it before you measure again.

  2. Then take a new temperature.

Frequently Asked Questions

My ThermoScan 3 still shows Err. What next?

Braun says to make sure the probe tip and lens are clean, with a new, clean disposable lens filter attached, and that the thermometer is inserted properly.
Then take a new temperature.
If it still shows Err, contact Braun's Consumer Relations team.

Does Err mean something else on the No Touch + Forehead thermometer (NTF3000)?

Yes: Braun gives it two meanings there.
The measured temperature was lower than 34.4 °C (93.9 °F) or higher than 42.2 °C (108 °F): make sure the protective cap is removed, then measure again.
Or the room is outside 15 to 40 °C (59 to 104 °F): leave the thermometer in a room within that range for 30 minutes before you use it.

What if the display flashes instead?

Braun says a flashing display indicates a system error.
Wait 1 minute until the thermometer turns off by itself, then turn it on again.
If the problem persists, reset the thermometer by removing the batteries for a moment.