HI, LO
Braun Thermometer
Severity: MinorWhat it means
Braun's manual for the ThermoScan 7 (IRT 6520) and the IRT 6030 ear thermometers shows HI when the temperature taken is too high, and LO when it is too low.
Both mean the reading is outside the human temperature range of 34 to 42.2 °C (93.2 to 108 °F).
Braun gives one fix for both: a clean probe and lens, a new lens filter, and the thermometer inserted properly.
Affected Models
- Braun ThermoScan 7 (IRT 6520) and IRT 6030 ear thermometers
How to Fix It
-
Check that the probe tip and lens are clean.
-
Attach a new, clean lens filter.
-
Insert the thermometer properly in the ear, then take a new temperature.
If the person seems very unwell, contact a doctor whatever the thermometer shows.